Troubleshooting Eddy Current Array Probe Noise: Adjusting Centering Feet for Lift-Off Control

Eddy current array probes bring a lot to the table—faster inspections, increased coverage, and enhanced defect detection. However, even with all their advantages, array probes are not immune to lift-off variations. In fact, centering is even more critical for tubing array probes because inconsistent lift-off can significantly impact signal quality and data interpretation.


A key factor in eddy current testing is magnetic coupling efficiency, which must be carefully controlled. This efficiency is influenced by fill factor (in encircling coils and inside bobbin probes) and lift-off (in surface pancake coils). While pancake coils typically ride the surface in applications like pencil probes and conductivity probes, in tubing array probes, pancake coils do not ride the surface. Instead, the probe must be precisely centered to ensure the lift-off distance between each coil and the tube’s inner diameter remains as uniform as possible. Any variation in this distance can degrade signal quality, making defect detection less reliable.


Key System Checks Before Adjusting Centering Feet

If an inspection starts showing noisy or unstable data, it’s important to first verify key system components:


  • Instrument Configuration – Ensure that the correct frequency, gain, and filter settings are applied.

  • Cables & Adapters – Check for secure connections, damaged cables, or faulty adapters that could introduce noise.

  • Grounding & Shielding – Confirm that proper grounding techniques are in place to reduce electrical interference.

  • Probe Delivery Snorkel & Guide Tube Runs – Inspect for any inconsistencies in the inside diameter of the snorkel, especially if different sections or types of guide tubes have been connected together. Variations can cause excessive probe movement, affecting centering and lift-off stability.


Once these factors are ruled out, checking the centering feet is definitely one thing that needs to be considered, as improper centering can introduce lift-off variations that affect signal quality.


Why Does Lift-Off Matter?

Lift-off refers to the distance between the eddy current coil(s) and the test material surface. Variations in this distance can cause:


  • Fluctuating Signal Strength – Inconsistent lift-off leads to unstable amplitudes, making defect sizing unreliable.

  • Loss of Sensitivity – Excessive lift-off reduces signal strength, potentially masking small defects.

  • Unsteady Phase Response – Variations in lift-off can affect phase angles, complicating flaw characterization.


Because lift-off must be controlled for accurate inspection results, maintaining proper probe centering is essential.


When to Adjust Centering Feet

If data appears unstable, centering feet may need adjustment. Some indicators include:


  • Inconsistent Amplitudes Across Coils – Some coils registering lower or higher signals than expected.

  • Unexplained Lift-Off Variations – Data showing irregular responses in clean areas of the tube.

  • Erratic Background Noise – Fluctuating baseline signals unrelated to material changes or defects.

How to Adjust or Refresh Centering Feet for Optimal Lift-Off Control


  1. Remove and Inspect the Probe – Check for signs of wear or debris on the centering feet.

  2. Determine If the Feet Are Adjustable – Some centering feet are not adjustable in the traditional sense but are fixed in place. These feet may still be “refreshed” by carefully manipulating each foot outward using a flat tool, ensuring that each foot is expanded evenly to achieve proper probe centering.

  3. Verify Spring Tension or Mechanical Adjustment – If the probe uses spring-loaded feet, ensure they provide adequate pressure to keep the probe centered without excessive force.

  4. Check for Proper Tube Wall Contact – The feet should engage the tube wall consistently without excessive rocking, floating, or scraping.

  5. Test the Probe in a Known Good Tube – Run a short scan in a clean, defect-free section to assess whether adjustments have stabilized lift-off.

  6. Monitor Data Consistency – Compare pre- and post-adjustment scans to confirm improved signal stability.


Other Considerations in Lift-Off Control

While centering feet are critical, other factors can also contribute to lift-off-related issues, including:

  • Probe Cable Management – Excessive drag on the cable can introduce movement, affecting centering stability.

  • Tube Surface Condition – Corrosion, scale, or deposits can cause uneven lift-off effects.

  • Probe Wear – Older probes may have worn coil housings that introduce unwanted spacing variation.

  • Operator Pull Speed – Inconsistent withdrawal speed can create probe movement, influencing lift-off.


Conclusion & Call to Action

Eddy current array probes provide powerful inspection capabilities, but maintaining proper lift-off control is essential to getting reliable results. While checking the instrument configuration, cables, grounding, and guide tube consistency should be part of any troubleshooting process, centering feet should not be overlooked. Whether adjustable or fixed, ensuring proper centering by refreshing the feet as needed can minimize lift-off variations, improve signal consistency, and enhance defect detection accuracy.


For more troubleshooting tips, best practices, and expert insights on eddy current testing, visit EddyCurrent.com. Whether you're looking for training, equipment, or in-depth technical resources, EddyCurrent.com is the ultimate hub for everything related to eddy current testing. Stay ahead of the industry—explore, learn, and connect today!

 
 
 